Features: -Adopts the technology of wireless communication to carry out wireless control, this cast away the mess computer. -Design of die-block style: The main unit is separated with computer. The pressure measuring unit is controlled separately with the my electricity measuring unit, it pre-remains several digital interfaces in order to make the independence clinical utilization and upgrade easily. This design is also easy to maintain promptly, suitable for upgrading the software. -It has urine flow measuring function, it is sensitive and correct. When the apparatus has measured the first drop of urine, the equipment will start automatically. -The distinct traction system can adjust the point of view at all position, it is easy to operate and measure the urethral pressure. -The infusion unit is divided to the style of pushing and turning. -It adopts the perfusion style of formlessness to avoid the interference from other apparatus or signal. This can ensure the urethral pressure curve reflects the actual situation of the patient and the doctor can easy to diagnose. -The perfusion pump is controlled by computer. It will not be limited by current direction when the catheter is clamped. -All the measurements conform to the standards of the International Continence Society (ICS). -It has been allocated with catheter both for adult(F9) and children(F6), so its clinical applicability is well. -The doctor can alter the set of parameter during the measurement. -The system is recommended by Urosource for basic Urodynamics.
Technical parameters: Measuring range of urine flow rate: 0ml/s~50ml/s, Error≤5% Measuring range of urine volume:10ml~1000ml, Error±2% Measuring range of pressure: -4.9kPa~ +19.6kPa(-50cmH2O~200cmH2O), Error≤5% Length of traction: ≥280mm Speed of traction: 0.5mm/s、1.0mm/s、2.0mm/s、4.0mm/s, Error≤5% EMG Range of measurement: 20μV~1mV CMRR: ≥80dB 20Hz~10kHz±20% Setting range of perfusion rate:2ml~ 80ml/min Range of perfusion volume: 10ml~1000ml, Error≤4% Analysis of pressure-rate, automatically generated four charts of (P-Q chart, A-G chart, LPURR chart, URA chart)
Measuring items: Urine Flow Rate, Vesical Function Test of Filling Period, Synchronic Test of Pressure-Uroflow-EMG, Analysis of Pressure / Flow Rate, Urethral Function Test, Analysis of Vesical Compliance, Analysis of Urine Leakage Point
Displaying curves: Curve of Vesical Pressure, Curve of Abdominal Pressure, Curve of Pdet, Curve of EMG, Curve of Urethral Pressure, Curve of Urethral Closure Pressure, Curve of Urine Flow Rate, Curve of Voided Volume
Compare List of SU-970 and Similar Product
|
SU-970 Series Product(our product) |
L Company |
Sensor |
Range |
-4.9kPa~+19.6kPa |
Same |
Uroflowmeter
Sensor |
Range/Error
Of Uroflowmeter |
0~50ml/s |
Same |
Range/Error
Of Urine Volume |
1000ml |
Same |
Tractor |
Length of Traction |
280 |
300mm |
Speed of Traction |
0.5~4mm/s |
0.5~5mm/s |
Perfusion |
Speed of Perfusion |
2~80ml/min |
1~150ml/min |
Other Property |
Self Limit Protection of Pressure Flow |
Same |
|
Style of Perfusion |
Combination of pushing pump to ensure the veracity of Urethra pressure measurement and Creep Rotation Perfusion Style |
Perfusion Style of Creep Rotation |
Myoelectricity |
Measuring Range |
20µF~1000µF |
5µF~1000µF |
Type of Electrode |
Surface Electrode, Needle Electrode |
Same |
Parameter of Uroflow-rate measurement |
Maximum Urine Flow Rate,
Average Urine Flow Rate, Time of Urine Flow, Time of Reaching Peak Value, Time of Start Urination, Urine Flow Volume, Urine Flow rate, Curve Chart of Urine Volume |
Same |
Parameter of Pressure Measurement
|
Pressure of Bladder
Pressure of Urethra
Abdominal Pressure (pressure in rectum) |
Same |
Items of Synchronous Measurement |
Pressure of Bladder, Pressure of Urethra
Abdominal Pressure (pressure in rectum), Pressure of Detrusor, Urine Flow Rate, Urine Flow Volume, Myoelectricity |
Same |
Items of Urethra Function Measurement |
Urethra Pressure Measurement and Average Parameters |
Same |
Soft Ware Analysis Function |
Analysis of Pressure Flow Rate, Produced 4 charts automatically (P-Q Chart, A-G Chart, LPURR Chart, URA Chart) |
P-Q Chart, A-G Chart, LPURR Chart, URA Chart |
Analysis of Bladder Adaptability |
Same |
Analysis of Urine Leakage Point |
Same |
Embed parameter calibration soft ware |
Can do parameter calibration |
Incident marked automatically / amended by hand |
Same |
Automatic Zero Setting / Balance by Hand |
Automatic Zero Setting |
In Accordance with ICS standard |
Same |
Through data playback / amendment / save |
Data playback |
|